Troubleshooting: ACE spectrograph

From CCDSH
Jump to navigationJump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
  • Problem: ThAr calibration spectra seem to be very faint. Solution: the mirror directing the light to the calibration fiber in the MAXnet cabinet may be stuck. Restart the spectrograph a few times (acectrl --stop, acectrl --start).
  • Problem: bright features (tipically bright Ar lines above 6000 Å on the ThAr spectra) cause massive saturation in whole columns of pixels. Solution: this problem may be related to the erroneous behavior of the preflush sequence of the FLI camera on the spectrograph bench. A possible workaround is to read a few bias frames out after every image.
Fig. 1
  • Problem: Spectrum focus seems to be on 0 value. Solution: In case you remember the previous value of the spectrum focus, use the acectrl --spectrum-preset-focus value command. If you don't, the default value is 11600, and large deviations from this are highly unlikely, thus it might be used as well (also note that end-switches are installed).
  • Problem: Switching spectrograph mode does not work, only appears to. Solution: This can happen after a longer power failure. Stop and start the spectrograph a couple of times. It may take several tries.
  • Problem: The QSI guider camera stopped working, and gives the /usr/local/bin/qsictrl: connection to camera failed, aborting. error message. Also, the camera probably beeps loudly, and the led on its back glows red. Solution: manual reset (unplug the camera and plug it in again). Be careful: the camera can easily be moved from the focal point by accident.
  • Problem: when trying to take an exposure ccdsh gives a sequence: imaging device is busy, waiting... or module: unable to initialize module 'mod_fli_camera.so':via initializer '__init': unable to access and/or initialize FLI CCD camera device. message. Solution: notify the support astronomer (or restart 172.31.171.98, Fig. 1., located in the RCC "observing room")
  • Problem: the spectrograph CCD cannot be cooled down or gives nonsense temperature values. Solution: the cooling system needs to be switched on. See section Warming-up of the camera on http://ccdsh.konkoly.hu/wiki/Spectrograph_CCD