Problem: ThAr calibration spectra seem to be very faint.

  • Solution: the mirror directing the light to the calibration fiber in the MAXnet cabinet may be stuck. Restart the spectrograph a few times (acectrl --stop, acectrl --start).
  • Solution: if it does not help, notify the support astronomer. A replacement for the ThAr lamp may be needed.

Problem: bright features (tipically bright Ar lines above 6000 Å on the ThAr spectra) cause massive saturation in whole columns of pixels.

  • Solution: this problem may be related to the erroneous behavior of the preflush sequence of the FLI camera on the spectrograph bench. A possible workaround is to read a few bias frames out after every image.
Fig. 1

Problem: Spectrum focus seems to be on 0 value.

  • Solution: In case you remember the previous value of the spectrum focus, use the acectrl --spectrum-preset-focus value command. If you don't, the default value is 11600, and large deviations from this are highly unlikely, thus it might be used as well (also note that end-switches are installed).

Problem: Switching spectrograph mode does not work, only appears to.

  • Solution: This can happen after a longer power failure. Stop and start the spectrograph a couple of times. It may take several tries.

Problem: The QSI guider camera stopped working, and gives the /usr/local/bin/qsictrl: connection to camera failed, aborting. error message.

Also, the camera probably beeps loudly, and the led on its back glows red.

  • Solution: manual reset (unplug the camera and plug it in again). Be careful: the camera can easily be moved from the focal point by accident.

Problem: when trying to take an exposure ccdsh gives a sequence: imaging device is busy, waiting... or module: unable to initialize module 'mod_fli_camera.so':via initializer '__init': unable to access and/or initialize FLI CCD camera device. message.

  • Solution: notify the support astronomer (or restart 172.31.171.98, Fig. 1., located in the RCC "observing room")

Problem: the spectrograph CCD cannot be cooled down or gives nonsense temperature values.